|
|
|
Gas Detection
Calibration Key to Accurate Readings |
Gas monitoring instruments are designed to protect personnel from
unseen hazards that may exist in workplace environments, including
confined spaces. It is vital to worker safety that these instruments
are maintained and calibrated properly.
Instrument inaccuracy due to improper or irregular calibration can
lead to serious accidents. Exposure to excessive levels of toxic gas
or an oxygen-deficient environment can cause workers serious illness
and even death. Combustible gas explosions are often catastrophic,
injuring or killing personnel and destroying property.
The only way to guarantee that an instrument will detect gas
accurately and reliably is to test it with a known concentration of
gas, says the International Safety Equipment Association (ISEA). The
ISEA recommends, at a minimum, verification of sensor accuracy
before each day’s use. Exposing
the instrument to a known concentration of test gas will show
whether the sensors respond
accurately and whether the instrument alarms function properly.
Calibration: The Key
to Accurate Readings
Gas detection instruments are used to detect the presence of toxic
and combustible gases, as well as oxygen deficiency or oxygen
enrichment (a fire and explosion hazard). Workers cannot rely on
their sense of smell to alert them to odorless hazards,
necessitating the use of gas detectors whenever a worker enters an
area with potential atmospheric hazards.
“Calibration” refers to an instrument’s measuring accuracy relative
to a known concentration of gas. Gas detectors measure the
concentration of a gas in an air sample by comparing the sensor’s
response to the response generated by a calibration gas of a known
concentration. The instrument’s
response to the calibration gas serves as the measurement scale or
reference point.
The responsiveness of electrochemical sensors will vary with
environmental conditions. Sensor response will be different (lower
or higher) depending on the actual environmental conditions.
Therefore, as much as possible, the monitors should be calibrated at
environmental conditions that
are the same as (or similar to) actual field conditions. Calibration
at locations where the equipment is to be used is always preferable.
Most instruments are equipped with two levels of alarms – warning
and danger. The warning alarm alerts the user that the environment
has a detectable concentration of gas and is therefore potentially
hazardous. The danger alarm indicates that the gas concentration
exceeds the programmed
“hazard” threshold, and the area is approaching a hazardous level.
Whether an instrument warns and/or alarms at the proper time depends
on its detection abilities and its ability to translate its findings
into an accurate reading.
Calibration Drift
If
the instrument’s reference point has shifted, the reading will shift
accordingly and be unreliable. This is called “calibration drift”
and it happens to all detectors over time. An instrument that
experiences calibration drift can still measure the quantity of gas
present but it cannot convert this information into an accurate
numerical reading. Regular calibration with a certified standard gas
concentration will update the instrument’s reference point, ensuring
that the instrument will produce continued, accurate readings.
Over time, the accuracy of gas detection instruments can diverge
from their calibration settings in several ways:
• Gradual chemical degradation of sensors and drift in electronic
components that occur naturally over time.
• Chronic exposures to, and use in, extreme environmental
conditions, such as high/low temperature and humidity, and high
levels of airborne particulates.
• Exposure to high (over-range) concentrations of the target gases
and vapors.
• Chronic or acute exposure of catalytic hotbead LEL sensors to
poisons and inhibitors.
These include: volatile silicones, hydride gases, halogenated
hydrocarbons, and sulfide gases.
• Chronic or acute exposure of electrochemical toxic gas sensors to
solvent vapors and highly corrosive gases.
• Harsh storage and operating conditions, such as when an instrument
is dropped onto a hard surface or submerged in liquid. Normal
handling/jostling of the equipment can create enough vibration or
shock over time to affect electronic components & circuitry.
Often, after exposure to the more extreme conditions above, when
calibration is attempted, the detector will either display a failure
message or it will not allow the user to fully adjust the display
reading. At this point, the severely damaged sensor must be replaced
and/or the detector serviced by
qualified personnel.
The primary reason for proper, regular instrument calibration is to
prevent inaccurate gas concentration readings that could lead to
injury or to death. Correctly calibrating an instrument helps to
ensure that the instrument will accurately respond to the gases that
it is designed to detect, warning
users of hazardous conditions before they reach dangerous levels.
In addition to detecting and correcting for calibration drift,
regular calibration assures the user that the instrument is
functional. Gas detection instruments are often subjected to harsh
operating and storage conditions where they can be damaged.
Both of these factors can affect instrument performance, leading to
inaccurate readings or even instrument failure. While a unit may
appear to be sound during a visual inspection, it actually could be
damaged internally. Regular calibration is the only way to be
certain that a detector is fully functional. Moreover, a standing
policy for regular calibration sets the tone for a safety-conscious
work environment and indicates to workers that safety is a priority.
As a result, workers may be more likely to keep safety principles in
mind throughout the workday.
A written record of calibration should be kept for the life of each
instrument. This record allows users to quickly identify an
instrument that has a history of excessive maintenance/repair or is
prone to erratic
readings.
Bump Tests vs.
Full Calibration
There are two methods of verifying instrument accuracy: a functional
or bump test and a full calibration, each appropriate under certain
conditions. A bump test verifies calibration by exposing the
instrument
to a known concentration of test gas.
The instrument reading is compared to the actual quantity of gas
present (as indicated on the cylinder). If the instrument’s response
is within an acceptable tolerance range of the actual concentration,
then its calibration is verified. (Note: It is recommended that
users check with the detection equipment manufacturer for the
acceptable tolerance ranges.) Instruments should be “zeroed” before
the bump test in order to give a more accurate picture of the bump
test results. When performing a bump test,
the test gas concentration should be high enough to trigger the
instrument alarm.
If the bump test results are not within the acceptable range, a full
calibration must be performed. A full calibration is the adjustment
of the instrument’s reading to coincide with a known concentration
(generally a certified standard) of test gas. For verification of
accuracy, calibration gas should
always be certified by and traceable to the National Institute of
Standards and Technology (NIST). In most cases, a full calibration
is only necessary when an instrument fails a bump test or after it
has been serviced. The full calibration and bump test should be
conducted in a clean
fresh air environment.
In the past, there often has been confusion regarding proper
calibration procedures and frequency. To clarify this issue, the
International Safety Equipment Association (ISEA) issued a position
statement on instrument calibration that states, “A bump test or
full calibration of direct-reading portable gas monitors should be
made before each day’s use in accordance with manufacturer’s
instructions,
using an appropriate test gas.” If the instrument fails a bump test,
it must be adjusted through a full calibration before it is used.
ISEA recommends more frequent testing if environmental conditions
that could affect instrument performance are suspected, such as
sensor poisons. The ISEA allows for less frequent calibration
verification under certain conditions (see below), but the interval
between testing should
never exceed 30 days.
According to the ISEA, less frequent verification may be appropriate
if the following criteria are met:
• During a period of initial use of at least 10 days in the intended
atmosphere, calibration is verified daily to ensure there is nothing
in the atmosphere to poison the sensor(s). The period of initial use
must be of sufficient duration to ensure that the sensors are
exposed to all conditions that might adversely affect the sensors.
• If the tests demonstrate that no adjustments are necessary, the
interval between checks may be lengthened, but it should not exceed
30 days.
• When calibrating an instrument, always follow the instrument
user’s manual for the manufacturer’s recommended calibration
frequency and procedure.
Calibration
Rules
The following are a few basic instrument calibration rules to ensure
a clear path to health and safety.
• Follow the manufacturer’s guidelines for proper calibration. No
job, including instrument calibration, can be performed properly or
safely without the right tools. The type and concentration
of calibration gas, sample tubing, flow regulators and calibration
adapters are key links in the calibration chain. Using equipment
provided by the original manufacturer should ensure a proper
start to every calibration.
• Only use certified calibration gas before its expiration date. The
most important tool used in calibration is the gas itself. The
instrument can only be as accurate as the gas used to calibrate it.
Be certain your supplier can provide a traceable certificate of
analysis for every calibration gas cylinder.
The concentration of calibration gas, particularly the concentration
of reactive gases such as hydrogen sulfide or chlorine, will only
remain stable for a finite period of time. Never use calibration gas
after its expiration date.
• Train workers on the proper methods of calibration. Most
instruments are designed to be field calibrated with instructions
detailed in user manuals, training videos or computer-based training
modules. Everyone responsible for performing instrument calibration
should be trained and tested
accordingly.
There is an inherent risk in many workplaces of injury or illness
from respiratory hazards such as oxygen deficiency and the presence
of toxic gases. Detection technology and products exist to minimize
such risk. Properly verifying the accuracy of gas detection
equipment before each day’s use
will help to ensure that each worker finishes the job safely.
FSM
SOURCE: Safety and Health Information Bulletin; U. S.
Dept. of Labor’s Occupational Safety and Health Administration,
Directorate of Science, Technology and Medicine, Office of Science
and Technology Assessment.
|
|
|