|
Calibration Basics for Gas
Detection Equipment
B Y
MICHAEL
D. SHAW
It is quite unlikely
that you will
ever use an absolute method for gas detection. Rather, you will employ
any one of dozens of “relative” methods — that is, methods that produce
some output that must be calibrated against a known standard. Then, its
display can be directly read out in units of concentration, usually
parts-per-million (ppm).
Even though proper calibration is 90 percent of
successful gas detection, it is a subject that has been neglected–often
purposely – by the majority of instrument manufacturers.
There’s a good reason for this, of course: Proper
calibration can often be difficult and expensive.
Early occupational health, toxic gas detection focused
on carbon monoxide (CO) and hydrogen sulfide (H 2S).
The calibration standards were supplied as gas blends in cylinders, and
in the case of CO, at least, things worked out pretty well. This is
because CO is not very reactive, and, within reason, maintains a stable
concentration in the cylinder, as the pressure drops with use.
On the other hand, H2S
is quite reactive, and the original simplistic techniques used to create
the cylinder gas blends could not provide a stable product. The problems
observed with H2S
blends were soon seen in blends for many other toxics. To make matters
worse, improper analogies were drawn between experiences in combustible
gas detection and toxic gas detection, establishing a false sense of
security about poorly prepared gas blends.
In fact, other than the obvious point that both
combustible and toxic gas detection get involved with detecting gases,
the two fields of endeavor could not be more different.
The combustible gases of interest are nearly all stable
(unless they are ignited by some external source), while nearly all
toxic gases are unstable, and in many cases are extremely reactive. Most
importantly, though, combustible gas detection is done in percent level
concentrations, while toxic gas detection is done in parts-per million,
and even parts-per-billion concentrations, 10,000 and 10 million times
lower, respectively.
Fortunately, calibration gas blending technology has
improved, encompassing specialized techniques for passivating the
cylinders, as well as logging experience to determine how long a blend
must age to become stable, and how long stability can be guaranteed.
Much of the technological development has been done with aluminum
cylinders, since this material seems to be less prone to wall effects
and unwanted chemical reactions than steel.
Here are some helpful hints:
• Order the blend so that the
concentration is about 50 percent of the instrument’s measuring range;
• Ensure that the blend’s analysis is
± 2 percent accurate (or better);
• Insist on NIST-traceability; and
• Obtain a written guarantee as to
how long the blend will be stable.
Since most of the cost of the blend
is in the analysis labor, order the largest cylinder you can use. Stay
away from disposable cylinders, which just become a solid waste problem.
After all, we ARE in the environmental business.
Permeation Devices
Certain toxic gases, especially at
low concentrations, are not well suited to being stored in cylinders,
and cylinder blends are cumbersome to re-standardize, in that a separate
(usually wet chemical) analytical method is required. An alternative is
permeation devices.
Permeation devices are small, inert
capsules containing a pure chemical compound in a two-phase equilibrium
between its gas phase and its liquid or solid phase. At a constant
temperature, the device emits the compound through its permeable portion
at a constant rate. This rate can always be determined via differential
weighing at constant temperature. The final concentration is adjusted by
varying the flow rate of an inert carrier gas.
Permeation devices are typically
employed within specially made systems that offer temperature and
carrier gas flow rate control.
Zero Gas
As you can imagine, if your
measurement range is in the low ppm (or less), accurately zeroing the
instrument is of vital importance. Consider that it is not a trivial
matter to remove contaminants such as carbon monoxide from air below
tenths of a ppm.
Zero air can be obtained from the same vendors who manufacture gas blends. We would make the following recommendations:
• Tell your supplier your target gas and measuring range, and have him suggest the proper zero gas for your application; and
• Ask for a written analysis of the zero gas.
Ideally, there will be specific information and not just a series of “less thans.” In general, the lower your measuring range, and the greater accuracy you desire, then the more frequently you should calibrate. Calibration monthly is a good median recommendation, and bi-monthly is even better.
When we say “calibration,” we mean a good patient effort that allows for sensor and instrument stabilization, to get a solid, reproducible reading. So-called bump tests, that challenge the instrument with some unknown, but high concentration of gas prove little, and can often be misleading. For the most part, these are NOT recommended.
In certain cases, less frequent calibration will still afford satisfactory results. This should be discussed with your instrumentation supplier.
FSM
Michael D. Shaw
is executive vice president and director of marketing for Interscan Corp., a manufacturer of toxic gas detection instrumentation and related software.
|
|
|