“Accuracy” as defined in the ANSI/ISA
51.1-1979 (R1993) standard Process Instrumentation
Terminology is: “The degree
of conformity of an indicated value
to a recognized accepted standard value,
or ideal value.”
Two common methods of rating or expressing
accuracy are:
• As a percent of scale length (or percent
of full scale). This rating method is
most commonly employed with instruments
equipped with an analog meter.
• As a percent of actual output reading.
This rating method has become more
popular for instruments provided with a
digital meter.
The above referenced standard also
notes that:
As a performance specification, accuracy
(or reference accuracy) shall be
assumed to mean accuracy rating of the device, when used at reference operating
conditions. Accuracy rating includes
the combined effects of conformity,
hysteresis, dead band, and repeatability
errors.
What Accuracy Is Not
In the world of gas detection instrumentation,
as in other places, confusion
exists. Consensus standards are helpful, of
course, but cannot be expected to address
all practical issues.
Accuracy is Not Minimum Detectability
“Minimum detectability” is simply the
lowest meter reading or other type of instrument
output that can be unambiguously
discriminated from noise. Some
agencies set a standard that minimum detectability
must be at least 2-2.5 times the
noise level.
Note that any data garnered at the level
of minimum detectability will not be very
accurate. For example, in a typical case,
the minimum detectability of a particular
instrument, provided with an analog meter,
is given as 1 percent of full scale, and accuracy is ± 2 percent of full scale.
Thus, for a 0-100 ppm scale, the minimum
detectable reading of 1 ppm would
actually be 1 ppm ± 2 ppm—hardly a
useful measurement.
Similarly, on a digital unit, the minimum
detectability of a particular instrument
is often given as the least significant
digit. On a commonly used 31/2 digit meter,
for a range of 0-199.9 ppm, this would
be 0.1 ppm. In this case, accuracy is specified
at ± 2 percent of reading ± 1 least
significant digit. Here, the minimum detectable
reading of 0.1 ppm would actually
be 0.1 ppm ± 0.002 ppm ± 0.1 ppm.
While technically better than the analog
example, this is still of little value.
Even so, knowing the minimum detectability
of an instrument can be helpful in situations when “go/no-go” readings
are of interest. Given a properly calibrated
instrument, the smallest observable response
would be—by definition—the
minimum detectable level, and would indicate
at least the presence of the substance
in question (any interferences
notwithstanding).
Of course, such practices should only
be done when instruments with more
appropriate sensitivity are not available.
Accuracy is Not Precision
“Precision” as defined in ASTM Standard
D 1356-05 Standard Terminology
Relating to Sampling and Analysis of Atmospheres
is:
The degree of agreement of repeated
measurements of the same property, expressed
in terms of dispersion of test results
about the mean result obtained by repetitive
testing of a homogeneous sample under
specified conditions.
A classic illustration shows a target
with many arrows closely clustered
around the bull’s-eye. This scenario
would be both accurate and precise. If many arrows were closely clustered far
from the bull’s-eye, the situation would
be precise, but not accurate. If arrows
were all over the target, those closest to
the bull’s-eye would be accurate, but the
archery session was not precise.
However, while this understanding of
precision holds true in the fields of science,
engineering, and statistics, as the
term is used in the world of computing,
precision can be either:
The number of significant decimal digits
or bits by which a particular value is
expressed; for example, a calculation
which rounds to three digits is said to have
a working precision or rounding precision
of 3.
Or…The units of the least significant
digit of a measurement; for example, if a
measurement is 25.371 meters then its
precision is millimeters (one unit in the
last place, or “ulp,” is 1 mm).
Accuracy is Not Resolution
“Resolution” as defined in the
ANSI/ISA-51.1-1979 (R1993) standard
Process Instrumentation Terminology is: The least interval between two adjacent
discrete details which can be distinguished
one from the other.
We can invoke another use of “resolution”
to help visualize this concept. In the
computer world, “resolution” is a measure
of the sharpness of an image or of the
fineness with which a device (such as a
video display, printer, or scanner) can produce
or record such an image. Resolution
in this context is usually expressed as the
total number or density of pixels in the
image—typically as dots per inch or dots
per millimeter.
For example, a 600-dpi (dots per inch)
printer is one that is capable of printing 600
distinct dots in a line 1 inch long. Thus, it
can print 360,000 dots per square inch.
Now, draw the analogy between all
these “distinct dots” and the least significant
digit on a digital meter. We referred
earlier to a 31/2 digit meter, set up for a measuring
range of 0-199.9 ppm. Our least significant
digit here is 0.1 ppm. There is no
way to read a digital meter beyond this
least significant digit. Or, to put it another
way, the resolution of the meter is 0.1 ppm.